Key Features:
- Vortex vacuum pumps ensure continuous 24/7 operation and are available in single or duplex (redundant) configurations.
- Includes a check valve and filtration system.
- Complies with GOST ISO 8835-3-2012.
- Extended alarm notifications.
- It can be equipped with a remote control panel and an adsorber for collecting nitrous oxide or other anaesthetic gases during anaesthesia or sedation.
Technical specifications:
| Model | Capacity, L/min | Capacity, Nm³/h | Max Vacuum, mbar | Blower Model | Power, kW | Power Supply |
| SEAG-145-O | 1×2410 | 1×145 | –160 | 2GH 410-H16 | 1×0.85 | 380 V, 50 Hz |
| SEAG-145-D | 2×2410 | 2×145 | –160 | 2GH 410-H16 | 2×0.85 | |
| SEAG-210-O | 1×3500 | 1×210 | –170 | 2GH 510-H16 | 1×1.3 | |
| SEAG-210-D | 2×3500 | 2×210 | –170 | 2GH 510-H16 | 2×1.3 | |
| SEAG-345-O | 1×5750 | 1×345 | –200 | 2GH 630-H16 | 1×2.2 | |
| SEAG-345-D | 2×5750 | 2×345 | –200 | 2GH 630-H16 | 2×2.2 | |
| SEAG-416-O | 1×6930 | 1×416 | –220 | 2GH 730-H26 | 1×3 | |
| SEAG-416-D | 2×6930 | 2×416 | –220 | 2GH 730-H26 | 2×3 | |
| SEAG-UP | Remote control panel for anaesthetic gas evacuation system | |||||
Note: Pumps are oil-free
SEAG-145-O — anaesthetic gas evacuation system, 145 — capacity in Nm³/h, O/D — single/duplex.
Control Panel:
The system features an intuitive control panel that enables:/p>

- Switching between vacuum pumps;
- Remote control;
- Error notifications;
- Maintenance alerts.